Copyright 2024 by Rex Instruments
 
 
| Cancel
ProComSol develops HART communicators for mobile device software that bridges the gap between HART Protocol devices, allowing for configuration, monitoring, calibrating, and troubleshooting of HART instruments in industrial plants.
Mobile :: Brands Masthead
Tradinco Traqc-20 LPC Low Pressure Calibrator

Tradinco Traqc-20 LPC Low Pressure Calibrator

Product Description

 
 
 
   
 
The LPC measures and generates pressures and consequently it is more than a mere measuring instrument – it can also be applied as test and calibration instrument for pressure sensors and pressure switches. Thanks to internal pressure generation no auxiliary tools are required for operation. An internal Li-ion battery is available for mobile operation. Parallel operation of mains and battery supply allows for maximum flexibility.
 
The Traqc-20 LPC is particularly suitable for stable pressure control in the very low pressure ranges (<<1000mbar), which is often used when environmental sensors need to be calibrated for cleanrooms, hospital areas and high-quality filters.
 
  • Very low pressure calibrator 0...1 to 0...1000 mBar
  • High precision
  • Compact size
  • Easily operated
  • Li-ion battery
 
 
Specifications
  • Pressure range 1 mbar, 10 mbar, 100 mbar, 1000 mbar
  • Accuracy 0.3 % FS, 0.1 % FS, 0.1 % FS, 0.1 % FS
  • Type of pressure Positive or negative gauge pressure or differential pressure
  • Pressure units Pa, kPa, hPa, bar, mbar, psi, inH2O, inHg, mmHg, Torr, mmH2O
  • Adjustment accuracy 0.05 %
  • Long-term stability < +/- 0.1% FS/year
  • Temperature drift ≤ 0.03 % FS/K 0.0 % with automatic of manual zero-point adjustment
  • Usable pressure range -10 - 110 %
  • Zero balance Automatic: in adjustable time intervals Manual: ZERO button
  • Instruments version desktop instrument with carry handle
  • Dimensions 102.6 x 257 x 271 mm without carry handle
  • Weight 4.5 kg
  • Ingress protection IP20
  • Pressure generation internal, electric pump
  • Screen Graphic display
  • Resolution 5 digits
  • Keyboard membrane keypad Menu
  • language German, English, Spanish and French
  • Connections Pressure connections 6.6 x 11 (hose diameter, D = 6 mm)
  • Power supply DC 24 V, 1 A
  • Battery type Li-Ion
  • Battery life approx. 8 h
  • Media Ambient air
  • Operating temperature 10 ... 40 °C
  • Storage temperature -10 ... 70 °C Specifications Di mensions
 
 
 
 
 
 
Copyright 2024 by Rex Instruments
Google Site Search